You are:
The INFINEED lt;precision plus lt; 3-way DEHP free, BPA free PVC extension has been designed to minimise its environmental impact, using only raw materials that are free from endocrine disruptors (elimination of components, including DEHP or BPA). The extension has been tested at pressures well above those recommended for gravity infusion; laboratory tests have shown a pressure resistance of approximately 10 bar. Contact
with 0.5% or 2% Chlorhexidine in 70° Isopropyl alcohol, without evaporation
time, and while the luer lock is connected, therefore under physical stress,
presents a significant risk of cracking of the luer lock, and should be avoided
as much as possible.
Extension with:
BPA-free 3-way stopcock: Lipid-resistant (96 hours to 7 days) with large valve stopcock and arrows indicating the direction of flow, permanently attached to the extension tube.
Tubing: PVC (DINCH) designed for comfort, optimal flow adjustment consistency and to prevent kinks. Diameter: 2.5 x 4.1mm
End cap: Mobile Luer Lock
End stopper: 0.8 µm purge filter with hydrophobic membrane
Resistance to pressure of 2 bar according to the claim of compliance with reference standard NF EN ISO 8536-9, 10 bar according to tests performed by Didactic (see scientific and technical file: Lipid-resistant from 96 hours to 7 days)
Made in China.
In accordance with the recommendations of the SF2H (2019), disinfecting the
connectors (when necessary) with 70° ethyl alcohol is always preferable, as it
is just as effective in disinfecting inert materials and there is much less
risk of altering plastic materials.
Indications
This medical device is intended for the parenteral administration of injectable preparations (drugs, nutrition).
Dispose according to regulation in member state.
Used materials
PVC DINCH (DEHP<0.1%(m/>)): Tubing - Polypropylene: Protective cap, stopcock stopper - Polyethylene: Stopcock valve - Tritan® BPA-free: Stopcock body - Acrylonitrile butadiene styrene: Fixed male Luer Lock, fixed female Luer lock
Labeling
on each unit are written the reference, the lot/serial number, the sterilization method and the expiry date if applicable
medical device
sterile
conformity
economic operator